MB-Rx In-Situ Reaction Monitoring Analyzer

Brand: ABB Group

The MB-Rx is designed around the key concepts of analytical performance, reliability and simplicity. It is a plug-and-play solution that can be used at any time without requiring any consumables or maintenance!

The MB-Rx features a number of integrated software modules that include all of the functions required for collecting and analyzing real-time data generated during chemical experiments. The Reaction Configuration Wizard takes the user through the critical steps of the experiment set-up process. The system can generate multiple real-time trends related to spectral features or chemical properties. Experiments can be modified “on the fly” using the full-featured main reaction interface. At the end of a reaction, the experiment data can be saved and reprocessed off-line using a variety of pre-treatments in order to optimize a reaction.

The MB-Rx features several integrated software modules that provide all functions to collect and analyze real-time data generated during chemical experiments. The user is taken through the critical steps of the experiment setup thanks to a reaction configuration wizard. The system can generate multiple real-time trends of spectral features or chemical properties. Experiments can be modified “on the fly” through the full-featured main reaction interface. At the end of a reaction, the experiment data can be saved and reprocessed off-line with different pretreatments, in order to optimize a reaction.
Features
  • Plug-and-play solution  
  • Maintenance free 
  • Continuous health monitoring  
  • Reduced footprint  
  • Upwards fiber launcher  
  • Non-hygroscopic optics  
  • Extremely stable double-pivot interferometer design  
  • Novel patented signal-processing algorithms  
  • Hastelloy insertion probe  
  • Rugged and liquid-tight fiber optic conduit  
  • Aluminum casting 
  • Choice of 3 types of ATR crystals  
  • Pre-aligned optics

Specifications

Spectral Range

  • ZnSe ATR probe + silver halide fibres: 600 to 3,300 cm-1
  • Diamond ATR probe + silver halide fibres: 600 to 1,900 cm-1 + 2,300 to 3,300 cm-1
  • ZrO2 ATR probe + chalcogenide fibres: 1,550 to 8,000 cm-1

Resolution

  • Apodized: adjustable 1 to 64 cm-1 (2r increments)

Signal Sampling

  • 24-bit ADC

Operating Temperature

  • -150° to 140° C

Detector

  • DTGS

Limit of Detection

  • For Acetone in Toluene: 0.1% w/w (60 s acquisition, 4 cm-1 resolution)

Enclosure

  • Casting: Rugged all-metal

Dimensions

  • 43.5 cm (W) x 28.0 cm (D) x 37.2 cm (H)

Weight

  • 27 kg

Power Supply

  • 120-240 Vac, 50/60 Hz

Power Consumption

  • 65 W

Humidity

  • 5% to 80%, relative humidity,  non-condensing

Certifications & Approvals

  • TUV
  • CE
show more